Vertical Laminar Flow Cabinets LORICA — LAMSYSTEMS
(Product protection)
Application:
Laminar flow cabinets are designed for creation of dust free abacterial air environment. They are used for work with low-risk substances and bacterial cultures, for protection of working materials from contaminated environment and for carrying out experiments in sterile working conditions.
The cabinets are designed only for protection of products harmless for personnel and the environment.
Model BAVnp-01-Laminar-S |
412.120 |
412.150 |
412.180 |
Width, m |
1.2 |
1.5 |
1.8 |
Dimensions of the assembled cabinet with the stand (WxDxH), mm |
1200x760x1870 |
1500х760х1870 |
1800х760х1870 |
Dimensions of the working chamber (WxDxH), mm |
1130x625x660 |
1430x625x660 |
1730x625x660 |
Mass of assembled cabinet with the stand, kg, not more than |
160 |
186 |
223 |
Class of the НЕРА-filters installed in the cabinet according to EN 1822-1 |
H14 |
H14 |
H14 |
Class of prefilter installed in the cabinet according to EN 779 |
G4 |
G4 |
G4 |
Air cleanliness class in the working chamber of the cabinet for suspended particles content according to ISO 14644-1 |
5ISO |
5ISO |
5ISO |
Average downflow velocity in the working chamber: – factory specified velocity, m/s – recommended range of velocity for independent adjustment whereby the uniformity of air flow is kept guaranteed, m/s |
0.40 0.25-0.50 |
0.40 0.25-0.50 |
0.40 0.25-0.50 |
Power consumption (without the built-in outlets load), W, not more than |
410 |
410 |
590 |
Noise level, dB, not more than |
55 |
55 |
55 |
Illuminance level in the working zone, lux, not less than |
1000 |
1000 |
1000 |
Standard delivery set incl. the support stand, UV-lamp.
Main features
- The fan motor control system SintelL-1 allows minimizing the power consumption of the cabinet, to reduce the level of acoustic and electromagnetic noise.
- The system of air consumption static regulation AIS LS automatically regulates air balance in the working chamber, changing the fan speed according to the level of filters contamination.
- The window rising mechanism allows fixing window in two positions which makes the working chamber and the window easy to clean from both sides with disinfectant solutions.
- HEPA filter is clamped with springs providing the filter tightness for the whole operation life.
- All electric control systems are placed outside of the main working chamber.
- The UV lamp is placed on the top of the rear side of the working chamber and protected by the stainless steel protective case.
- Capability of an operative adjustment the downflow velocity.
- Low noise level.
- Coloring of the cabinet is made with the powder enamel proof for disinfectant solutions.
- The cabinet is equipped with a hydraulic damper for smooth closing the front glass eliminating the possibility of its break and providing an easy operation.
- The laminarization unit provides uniform downward laminar flow.
- Access to the main filter from the front technical opening provides ease of the filter replacement.
Additional options:
– Gas tap
– Vacuum tap
– Extra sockets