Microbiological Safety Cabinets, Class II type A2 NEOTERIC
(Operator, environment and product protection)
Application
Microbiological safety cabinet class II NEOTERIC is produced in conformity with the requirements of European standard EN 12469:2000.
The cabinet provides product, operator and environment protection.
Model BMB-II-Laminar-S |
221.090 |
221.120 |
221.150 |
221.180 |
Width, m
|
0.9 |
1.2 |
1.5 |
1.8 |
Dimensions of the assembled cabinet with the stand (WxDxH), mm |
1000х770х2150 |
1200x770x2150 |
1500x770x2150 |
1800х770х2092 |
Dimensions of the working chamber (WxDxH), mm |
905x610x750 |
1105x610x750 |
1405x610x750 |
1705x610x750 |
Mass of assembled cabinet with the stand, kg, not more than |
196 |
235 |
300 |
300 |
Class of the НЕРА-filters installed in the cabinet according to EN 1822-1 |
H14 |
H14 |
H14 |
H14 |
Air cleanliness class in the working chamber of the cabinet for suspended particles content according to ISO 14644-1-2015, not less than |
5ISO |
5ISO |
5ISO |
5ISO |
Average velocity of the inflow though the work opening, m/s |
0.47 |
0.47 |
0.47 |
0.47 |
Average downflow velocity in the working chamber, m/s |
0.35 |
0.35 |
0.35 |
0.35 |
Power consumption (without the built-in outlets load), W, not more than |
180/100* |
180/110* |
330/160* |
590/150* |
Noise level**, dB, not more than |
59 |
55 |
59 |
59 |
Illuminance level in the working zone, lux, not less than |
1000 |
1000 |
1000 |
1000 |
Air recirculation rate in the cabinet, % |
≈70 |
≈70 |
≈70 |
≈70 |
Air volume supplied to the working chamber, m3/h |
656-674 |
795-817 |
1008-1036 |
1210-1245 |
Exhaust air volume, m3/h |
273-309 |
333-378 |
426-484 |
510-580 |
*The power consumption with newly installed uncontaminated HEPA filters.
Standard delivery set incl. the support stand, the pullout UV-unit, arm support, electromagnetic keys (3 pcs.).
Main features:
- EC fan motor control system SintelL-1 allows minimizing the power consumption of the cabinet and reducing level of acoustic and electromagnetic noise
- The system of air consumption static regulation AIS LS automatically regulates air balance in the work chamber changing the number of fan revolutions according to the level of filter contamination
- Control panel with an LC Display indicates switching on the cabinet, possible malfunctions, selected operational modes
- Sloped front sash widens the overview of the work chamber and the useful surface of the table top
- Cabinet’s housing is powder coated and resistant to disinfectant solutions
- The optical sensors of the front sash and the UV unit position provide total control of the cabinet operating modes
- Easy replacing of the filtering elements
- The table top is made of segments for a more convenient working zone disinfection and autoclaving
- A no-touch electromagnetic key effectively protects the control system from an unauthorized access.
A pull-out UV unit is an unique construction:
– is located outside the working chamber and doesn’t interrupt a laminar air flow
– doesn’t require a special place for storage
– reliable and easy-to-use
– well-adapted for cleaning
– completely covers the front opening of the cabinet within standby time
– is controlled by optical position sensor
Additional options:
Gas tap with an electromagnetic valve
Vacuum tap with an electromagnetic valve
Hood to connect the cabinet to an external exhaust system
LED light
ULPA filters
Extra sockets